What Gases Can a PID Sensor Detect? List of Detectable Gases for 10.6 eV Photoionization PID

classify:PID sensor   time:2026-11-04

Whether a PID sensor can detect a gas depends on one rule: the gas ionization potential must be lower than the energy of the UV lamp. The 10.6 eV lamp is the most widely used. 9.8 eV and 11.7 eV lamps are also available for special purposes.

Gases Detectable by PID Sensor

Aromatic hydrocarbons

Benzene, toluene, xylene, ethylbenzene, styrene, naphthalene

Alcohols

Methanol, ethanol, isopropanol, n-butanol

Ketones

Acetone, methyl ethyl ketone (MEK), cyclohexanone

Esters

Ethyl acetate, butyl acetate (common coating solvents)

Aldehydes

Formaldehyde, acetaldehyde (response varies by instrument)

Chlorinated solvents

Trichloroethylene, tetrachloroethylene, dichloromethane

Alkanes & alkenes

Hexane, heptane, cyclohexane, ethylene, propylene

Amines

Ammonia, dimethylamine, trimethylamine

A few inorganic gases

Hydrogen sulfide (H₂S), phosphine (limited response on 10.6 eV PID)

Important note: PID gives a total VOC reading. It cannot tell you which specific gas is present in the mixture.

Gases NOT Detectable by Standard 10.6 eV PID

Nitrogen, oxygen, CO₂, water vapor, methane, ethane, propane, carbon monoxide (CO), sulfur dioxide (SO₂), hydrogen cyanide (HCN), hydrogen chloride (HCl), freon, ozone, hydrogen.

Methane has an ionization potential of 12.61 eV. Standard PID cannot detect methane or natural gas leaks. For explosion hazard monitoring, use LEL catalytic bead sensors instead.

Extra Tips

  1. 9.8 eV lamp: narrower detection range, mainly used for benzene measurement;
  2. 11.7 eV high-energy lamp: covers more gases but shorter service life and higher cost;
  3. High humidity, high CO₂ and methane can cause signal quenching and baseline drift;
  4. PID only measures vapor, not liquid mist, aerosols or dust.