Overview of PID Sensor Working Principle
PID stands for Photoionization Detector. Its detection relies on the photoionization effect, which detects volatile organic compounds (VOCs) whose ionization potential is lower than the energy of UV photons. Most common PID sensors use a 10.6 eV UV lamp; 9.8 eV and 11.7 eV versions are available for special applications.
Complete Working Process
- Sample gas flows into the ionization chamber of the sensor;
- The UV lamp emits high-energy photons with fixed energy;
- VOC molecules absorb photons and get ionized into positive ions and free electrons if their ionization potential is lower than photon energy. Formula: \(\boldsymbol{RH + h\nu \rightarrow RH^+ + e^-}\);
- An electric field between collecting electrodes drives positive ions to the cathode and electrons to the anode, generating a tiny ion current;
- The weak current is amplified and filtered. The current value is proportional to the concentration of ionized VOC molecules;
- The circuit converts the electric signal into readable gas concentration in ppm or ppb.
Core feature: PID provides non-destructive measurement. Ions recombine into original molecules after leaving the electric field. The sample gas is not consumed or permanently changed.
Internal Components
- UV lamp: photon source; its energy determines detectable gas species;
- Ionization chamber: space where photoionization occurs;
- Collecting electrodes: capture ions and electrons to generate current;
- Signal conditioning circuit: amplify faint nA-level current, reduce noise and convert signals;
- Inlet filter membrane: block dust and mitigate humidity interference.
Pros and Cons from Its Working Principle
✅ Advantages: Rapid response, high sensitivity down to ppb level, perfect for on-site leak survey. Compact size for easy integration.❌ Limitations: Broad-spectrum response, cannot identify individual VOCs in gas mixtures. High humidity causes baseline drift. The UV lamp is a consumable with limited service life. Gases with higher ionization potential such as methane, CO₂, CO, N₂ and O₂ cannot be detected.
Typical Applications
Indoor TVOC air quality monitoring, chemical plant leak inspection, solvent monitoring in coating & printing workshops, soil and groundwater organic contamination screening, hazardous material emergency response.
PID Sensor Working Principle | How Photoionization PID VOC Sensor Operates
classify:PID sensor time:2026-11-04
BIKAI Technology Co., Ltd.